Buku la GDOT-80A Insulation Oil Tester-losinthidwa1105
●Chonde werengani bukuli mosamala musanagwiritse ntchito.
●Chonde onani ngati woyesayo alumikizidwa bwino padziko lapansi musanayese.
●Ndizoletsedwa kusuntha kapena kukweza chivundikiro choyesera poyesa kuti tipewe kuvulazidwa ndi mphamvu zambiri.Mphamvu ziyenera kuzimitsidwa musanalowe m'malo mwa mafuta a zitsanzo.
●Gwirani mosamala mukanyamuka kapena kutseka chivundikiro choyesera ma voltage apamwamba!
●Ngati choyesa chikugwira ntchito molakwika mafuta otsekereza atayika, chonde zimitsani choyesa kwa masekondi 10, ndikuyambiranso.
●Mapepala osindikiza akatha, chonde onani gawo lofotokozera chosindikizira (kapena chowonjezera pamanja) kuti musinthe pepala losindikiza kuti mupewe kuwonongeka kwa mutu wosindikiza.
●Sungani choyesa kutali ndi chinyezi, fumbi ndi zinthu zina zowononga, ndikuchisunga kutali ndi kutentha kwakukulu.
●Gwirani mosamala mumayendedwe.Osayika mmwamba mbali pansi.
●Bukhuli likhoza kukonzanso moyenerera popanda chidziwitso china pasadakhale.Ngati muli ndi funso, chonde titumizireni.
Nthawi ya chitsimikizo cha mndandandawu ndi chaka chimodzi kuyambira tsiku lotumizidwa.Chonde onani invoice yanu kapena zikalata zotumizira kuti mudziwe masiku oyenera a chitsimikizo.HVHIPOT corporation ikupereka chitsimikizo kwa wogula woyambirira kuti chinthuchi sichikhala ndi zolakwika pazakuthupi ndi kapangidwe kake pakagwiritsidwe ntchito wamba.Munthawi yonse yachitsimikizo, perekani kuti zolakwika zotere sizinatsimikizidwe ndi HVHIPOT kuti zidayamba chifukwa cha nkhanza, kugwiritsa ntchito molakwika, kusintha, kuyika molakwika, kunyalanyaza kapena kuwononga chilengedwe, HVHIPOT ili ndi malire pakukonza kapena kusintha chida ichi panthawi yachitsimikizo.
GDOT-80C chida | 1 pc |
Chikho chamafuta (250ml) | 1 pcs |
Chingwe champhamvu | 1 pc |
Fuse yapakati | 2 ma PC |
Ndodo yogwedeza | 2 ma PC |
Standard gauge (25mm) | 1 pc |
Sindikizani pepala | 2 mitu |
Tweezer | 1 pc |
Wogwiritsa ntchito | 1 pc |
Lipoti la mayeso a fakitale | 1 pc |
HV Hipot Electric Co., Ltd. yawerengera mosamalitsa komanso mosamala bukuli, koma sitingatsimikizire kuti palibe zolakwika ndi zosiyidwa kwathunthu m'bukuli.
HV Hipot Electric Co., Ltd. yadzipereka kupititsa patsogolo ntchito zogulitsa, ndikuwongolera magwiridwe antchito, kotero kampaniyo ikhalabe ndi ufulu wosintha zinthu zilizonse ndi mapulogalamu apulogalamu zomwe zafotokozedwa m'bukuli komanso zomwe zili m'bukuli popanda kuyambiranso. zindikirani.
Kusungunula kwamkati kwa zida zamagetsi zambiri zamakina amagetsi, masitima apamtunda, mafakitole akuluakulu a petrochemical ndi mabizinesi, nthawi zambiri amatengera mtundu wamafuta odzaza mafuta, chifukwa chake, kuyesa kwamphamvu kwamafuta a dielectric ndikofala komanso kofunikira.Kuti tikwaniritse zosowa za msika, tapanga ndi kupanga zoyesa zoyesa mphamvu zamafuta a dielectric molingana ndi muyezo wadziko lonse wa GB/T507-2002, muyezo wamakampani DL429.9-91 komanso makina aposachedwa a Electric Power Industry Standard DL/T846.7 -2004 tokha.Chida ichi, pogwiritsa ntchito kachipangizo kakang'ono kachipangizo kakang'ono kamene kamayambira, chimazindikira ntchito yodziwikiratu, kuyeza kolondola kwambiri, kuwongolera bwino ntchito ndikuchepetsa kulimbikira kwa ogwira ntchito.Komanso, ndi yaying'ono kukula kwake komanso yabwino kunyamula.
●Ndi microprocessor, kwaniritsani zokha kuyesa kwamagetsi kwa 0 ~ 80KV (kuphatikiza kulimbikitsa, kusamalira, kusakaniza, kuyimirira, kuwerengera, kusindikiza ndi zina).
●Chiwonetsero chachikulu cha LCD.
●Ntchito yosavuta.Makinawa amangomaliza kuyesa kuyesa kwamagetsi pa kapu imodzi yamafuta amafuta pambuyo pa kukhazikitsidwa kosavuta ndi woyendetsa.Mtengo wamagetsi owonongeka wa 1 ~ 6 nthawi ndi nthawi zozungulira zidzasungidwa zokha.Pambuyo pa mayesowo, chosindikizira chotenthetsera chidzasindikiza mtengo uliwonse wosweka ndi mtengo wapakati.
●Mphamvu-pansi kuteteza.Chithasungani zotsatira zoyesa 100 ndikuwonetsa kutentha ndi chinyezi.
●Adopt single-chip microcomputer kuti muwonjezere mphamvu yamagetsi pa liwiro lokhazikika.Ma frequency a voltage ndi olondola pa 50HZ, onetsetsani kuti njira yonseyo ndiyosavuta kuwongolera.
●Ndi over-voltage, over-current and limit protection kuonetsetsa chitetezo cha ogwira ntchito.
●Ndi ntchito yowonetsera kutentha koyezedwa ndi wotchi yamakina.
●Lumikizanani ndi kompyuta ndi mawonekedwe a RS232.
Magetsi | AC220V ± 10%, 50Hz |
Mphamvu yamagetsi | 0~80kv ku(Zosankhika) |
Mphamvu | 1.5kVA |
Mphamvu | 200W |
Kuthamanga kwamagetsi kumawonjezeka | 2.0~3.5k kuV/s (Zosintha) |
Votejikuyezakulondola | ±3% |
Kusokonezeka kwa Waveform | ≤3% |
Nthawi yowonjezera | 5min (Zosinthika) |
Nthawi yoyima | 15min (Zosintha) |
Nthawi zowonjezera | 1~6 (Zosankhika) |
Kuchitachilengedwe | Tkutentha: 0 ℃-45°C |
Hpakati:Max.chinyezi chachibale75% | |
Dimension | 465x385x425mm |
① chosindikizira chotentha--kusindikiza zotsatira zoyesa;
② LCD--kuwonetsa menyu, mwachangu ndi zotsatira zoyesa;
③ Makiyi ogwiritsira ntchito:
Dinani "◄" kiyi kuti muwonjezere mtengo;
Dinani "►" kiyi kuti muchepetse mtengo;
Sankhani--posankha ntchito (chinthu chosankhidwa chili pawonetsero);
Tsimikizirani--pakuchita ntchito;
Back--kwa exiting ntchito mawonekedwe;
④ Kusintha kwamphamvu ndi chizindikiro
1. Kukonzekera musanayesedwe
1.1 Lumikizani poyambira pansi (kumanja kwa zida) ku waya pansi mwamphamvu musanagwiritse ntchito zida.
1.2 Chotsani zitsanzo zamafuta molingana ndi muyezo woyenera.Sinthani mtunda wa elekitirodi mkati mwa kapu yamafuta molingana ndi geji yokhazikika.Yeretsani kapu molingana ndi zofunikira.Thirani chitsanzo cha mafuta mu kapu ndikutseka kapu.
1.3 Kusintha magetsi a AC220V zinthu zomwe zili pamwambazi zitatsimikiziridwa, zokonzekera kuyesedwa.
2. Kuyesedwa
2.1 Kanikizani chosinthira mphamvu ndiyeno lowetsani mawonekedwe otsatirawa:
2.2 Kusintha kwa Parameter System
Dinani batani la "Confirm" ndikulowa mu mawonekedwe awa:
Kupititsa patsogolo: ogwiritsa ntchito amatha kusankha malinga ndi zomwe akufuna.
Press "Back" kiyi kuti atuluke mawonekedwe awa pambuyo zoika zichitika.
2.3 Kuyesa
Dinani batani la "Sankhani" kuti musankhe "Start Test" menyu ndikusindikiza batani la "Tsimikizani" kuti mulowe mu mawonekedwe awa:
Kuti mupitilize kuyesa kotsatira mukangomaliza kuyesa koyamba mpaka ma frequency owonjezera akhazikitsidwa.Pomaliza, zotsatira zake zimawonetsedwa ndikusindikizidwa motere:
2.4 Kuwona ndi kusindikiza deta:
Dinani batani la "Sankhani" kuti musankhe menyu ya "Kuwonera ndi Kusindikiza" ndikusindikiza batani la "Tsimikizirani" kuti mulowe mawonekedwe otsatirawa:
Sankhani "Tsamba Pamwamba" kapena "Tsamba Pansi" ndikusankha zolemba zomwe zisindikizidwe ndikusankha "Sindikizani".
●Kusankhidwa kwa zitsanzo zamafuta ndi kuyika kwa mtunda wa electrode kudzakwaniritsa zofunikira zadziko ndi mafakitale.
●Ogwiritsa ntchito kapena antchito ena amaletsedwa kukhudza chipolopolocho mphamvu ikayatsidwa kuti apewe ngozi.
●Mphamvuyo idzadulidwa nthawi yomweyo ngati vuto lililonse lachilendo likupezeka panthawi ya opaleshoni.
●Chida ichi sichidzawonetsedwa m'malo achinyezi.
●Sungani kapu yamafuta ndi maelekitirodi oyera.Lembanikapu yokhala ndi mafuta a transfoma atsopano kuti atetezedwe ikapanda ntchito.Yang'anani mtunda wa electrode ndikuyang'ana kulimba pakati pa nsonga ya electrode ndi ulusi wa electrode bar screw musanagwiritse ntchito chikho.
1. Mafuta Cup Kuyeretsa Njira
1.1 Pukuta ma elekitirodi ndi zitsulo mobwerezabwereza ndi nsalu yoyera ya silika.
1.2 Sinthani mtunda wa electrode ndi geji yokhazikika
1.3 Gwiritsani ntchito petroleum ether (zosungunulira zina za organic ndizoletsedwa) kuyeretsa katatu.Nthawi zonse amatsatira ndondomeko zotsatirazi:
① Thirani mafuta a petroleum ether mu kapu yamafuta mpaka kapu itadzaza 1/4 ~ 1/3.
② Phimbani m'mphepete mwa chikho ndi galasi lotsukidwa ndi petroleum ether.Gwirani chikhocho mofanana kwa mphindi imodzi ndi mphamvu zina.
③ Thirani mafuta a petroleum ether ndikuwumitsa kapu ndi chowuzira kwa mphindi 2-3.
1.4 Gwiritsani ntchito chitsanzo cha mafuta kuti muyesedwe kuti muyeretse chikhomo kwa nthawi 1-3.
① Thirani mafuta a petroleum ether mu kapu yamafuta mpaka kapu itadzaza 1/4 ~ 1/3.
② Phimbani m'mphepete mwa chikho ndi galasi lotsukidwa ndi petroleum ether.Gwirani chikhocho mofanana kwa mphindi imodzi ndi mphamvu zina.
③ Thirani mafuta akumanzere ndikuyesa kumayamba.
2. Njira Yoyeretsera Ndodo
2.1 Pukuta ndodo yogwedeza mobwerezabwereza ndi nsalu yoyera ya silika mpaka tinthu tating'ono tating'ono tisapezeke pa malo awo.Ndizoletsedwa kukhudza malo ndi manja.
2.2 Gwiritsani ntchito zokakamiza kuti mutseke ndodo;ikani mu petroleum ether ndikutsuka.
2.3 Gwiritsani ntchito mphamvu zomangira ndodo ndikuyimitsa ndi chowuzira.
2.4 Gwiritsani ntchito zokakamiza kuti mutseke ndodo;Ikani izo mu chitsanzo mafuta ndi kusamba.
3. Mafuta Cup yosungirako
Njira 1 Dzazani chikhocho ndi mafuta abwino otsekera mayeso atatha ndikuyiyika mokhazikika.
Njira 2 Yeretsani ndi kupukuta chikhocho motsatira ndondomeko zomwe tatchulazi ndikuziyika mu chowumitsira chowumitsira.
Zindikirani: Chikho cha mafuta ndi ndodo yogwedeza idzatsukidwa pansi pa ndondomeko pamwambapa mutatha kuyesa koyamba ndikuyesa ndi mafuta osauka.
4. Common Fault Clearrances
4.1 Kuzimitsa magetsi, palibe zowonetsera pazenera
① Onani kuti pulagi yamagetsi ndiyolimba kapena ayi.
② Onani fuse yomwe ili mkati mwamagetsi ili bwino kapena ayi.
③ Yang'anani magetsi a soketi.
4.2 Chikho chamafuta popanda chodabwitsa
① Onani kuyika kwa zolumikizira pa board board.
② Yang'anani kukhudzana ndi chosinthira champhamvu kwambiri pachivundikiro chamilandu.
③ Yang'anani momwe zimalumikizirana ndi ma voltage apamwamba.
④ Yang'anani kuphulika kwa chingwe champhamvu kwambiri.
4.3 Kusiyanitsa kwa chiwonetsero sikukwanira
Sinthani potentiometer pa bolodi lozungulira.
4.4 Kulephera kwa Printer
① Onani pulagi yamagetsi ya chosindikizira.
② Onani kulumikizidwa kwa mzere wa data wa printer.