GDW-106 Mafuta Dew Point Tester
●Malangizo otsatirawa amagwiritsidwa ntchito ndi munthu woyenerera kuti apewe kugwedezeka kwamagetsi.Osachita ntchito iliyonse yopitilira malangizo ogwirira ntchito pokhapokha ngati muli oyenerera kutero.
●Osagwiritsa ntchito chipangizochi pamalo oyaka komanso pachinyontho.Sungani pamwamba paukhondo ndi youma.
●Chonde onetsetsani kuti zida zili zowongoka musanatsegule.Osagwetsa zida kwambiri pewani kuwonongeka kwa zida.
●Ikani zidazo pamalo owuma, aukhondo, ndi mpweya wabwino wopanda mpweya wowononga.Kuyika zida popanda zotengera zoyendera ndizowopsa.
●Panena iyenera kukhala yowongoka panthawi yosungira.Kwezani zinthu zosungidwa kuti muteteze ku chinyezi.
●Osasokoneza chida popanda chilolezo, zomwe zingakhudze chitsimikizo cha mankhwala.Fakitale ilibe udindo wodzigwetsa yokha.
Nthawi ya chitsimikizo cha mndandandawu ndi chaka CHIMODZI kuyambira tsiku lotumizidwa, chonde onani invoice yanu kapena zikalata zotumizira kuti mudziwe masiku oyenera a chitsimikizo.HVHIPOT corporation ikupereka chitsimikizo kwa wogula woyambirira kuti chinthuchi sichikhala ndi zolakwika pazakuthupi ndi kapangidwe kake pakagwiritsidwe ntchito bwino.Munthawi yonse yachitsimikizo, perekani kuti zolakwika zotere sizinatsimikizidwe ndi HVHIPOT kuti zidayamba chifukwa cha nkhanza, kugwiritsa ntchito molakwika, kusintha, kuyika molakwika, kunyalanyaza kapena kuwononga chilengedwe, HVHIPOT ili ndi malire pakukonza kapena kusintha chida ichi panthawi yachitsimikizo.
Ayi. | Dzina | Qty. | Chigawo |
1 | GDW-106 Host | 1 | chidutswa |
2 | Electrolytic cell botolo | 1 | chidutswa |
3 | Electrolytic electrode | 1 | chidutswa |
4 | Kuyeza electrode | 1 | chidutswa |
5 | Electrolytic cell jakisoni pulagi | 1 | chidutswa |
6 | Magalasi aakulu akupera pulagi | 1 | chidutswa |
7 | Pulagi yaing'ono yamagalasi (notch) | 1 | chidutswa |
8 | Small galasi akupera pulagi | 1 | chidutswa |
9 | Ndodo yogwedeza | 2 | ma PC |
10 | Ma gel osakaniza a silika | 1 | thumba |
11 | silika gel osakaniza | 9 | ma PC |
12 | 0.5μl micro sampler | 1 | chidutswa |
13 | 50μl micro sampler | 1 | chidutswa |
14 | 1 ml ya micro sampler | 1 | chidutswa |
15 | Molunjika youma chubu | 1 | chidutswa |
16 | Chingwe champhamvu | 1 | chidutswa |
17 | Vacuum mafuta | 1 | chidutswa |
18 | Electrolyte | 1 | Botolo |
19 | Sindikizani pepala | 1 | gudubuza |
20 | Wogwiritsa ntchito | 1 | chidutswa |
21 | Lipoti la mayeso | 1 | chidutswa |
HV Hipot Electric Co., Ltd. yawerengera mosamalitsa komanso mosamala bukuli, koma sitingatsimikizire kuti palibe zolakwika ndi zosiyidwa kwathunthu m'bukuli.
HV Hipot Electric Co., Ltd. yadzipereka kupititsa patsogolo ntchito zogulitsa, ndikuwongolera magwiridwe antchito, motero kampaniyo ikhalabe ndi ufulu wosintha zinthu zilizonse ndi mapulogalamu apulogalamu omwe afotokozedwa m'bukuli komanso zomwe zili m'bukuli popanda kuyambiranso. zindikirani.
Ukadaulo wa Coulometric Karl Fischer umagwiritsidwa ntchito kuyeza ndendende chinyezi chomwe chili ndi zitsanzo.Ukadaulo umagwiritsidwa ntchito kwambiri pakulondola komanso mtengo wotsika mtengo woyesera.Mtundu wa GDW-106 umayezera ndendende chinyezi pamiyeso yamadzimadzi, olimba ndi mpweya malinga ndiukadaulo.Amagwiritsidwa ntchito mu magetsi, mafuta, mankhwala, zakudya ndi zina zotero.
Chidachi chimagwiritsa ntchito mayunitsi amphamvu opangira m'badwo watsopano komanso mawonekedwe atsopano ozungulira komanso kuti kugwiritsa ntchito mphamvu zochepa kwambiri kumapangitsa kuti chitha kugwiritsa ntchito batire yaying'ono yosungira komanso yonyamula.Kuweruza kumapeto kwa electrolysis kumatengera kuyesa chizindikiro cha electrode ndipo kukhazikika ndi kulondola ndizofunikira kwambiri pakutsimikiza kulondola.
●5-inch high-definition color touch screen, chiwonetserocho ndi chomveka komanso chosavuta kugwiritsa ntchito.
●Njira ziwiri zolipirira ma electrolyte opanda kanthu pakali pano komanso chipukuta misozi chowongolera kuti muwunikenso zotsatira zoyesa.
●Ntchito zowunika kuyeza ma elekitirodi otseguka chigawo ndi cholakwika chachifupi.
●Imatengera chosindikizira chaching'ono chotentha, kusindikiza ndikosavuta komanso mwachangu.
●Ma formula 5 owerengera amapangidwa mu chidacho, ndipo gawo lowerengera lazotsatira zoyeserera (mg / L, ppm%) lingasankhidwe momwe lingafunikire.
●Sungani zokha mbiri yakale ndi tabu yanthawi, yopitilira ma 500.
●Ma microprocessor amakono opanda kanthu amawongolera okha kubweza, ndipo ma reagents amatha kufika pamlingo mwachangu.
●Miyeso Yosiyanasiyana: 0ug-100mg;
●Kulondola muyeso:
Electrolysis madzi Kulondola
3ug-1000ug ≤±2ug
> 1000ug ≤± 02% (zigawo zomwe zili pamwambazi sizikuphatikiza cholakwika cha jakisoni)
●Kusamvana: 0.1ug;
●Electrolyzing Current: 0-400mA;
●Kugwiritsa ntchito mphamvu zambiri: 20W;
●Kulowetsa Mphamvu: AC230V ± 20%, 50Hz ± 10%;
●Kugwira ntchito Kutentha kozungulira: 5℃ 40℃;
●Kugwiritsa Ntchito Chinyezi Chozungulira: ≤85%
●Kukula: 330 × 240 × 160mm
●Net Kulemera kwake: 6kg.
Kapangidwe ka Zida ndi Kusonkhana
1. Wolandira alendo
Chithunzi 4-1 Host
2. Electrolytic Cell
Chithunzi 4-2 Chithunzi cha kuwonongeka kwa ma cell a Electrolytic
Chithunzi 4-3 Chojambula chamsonkhano wamaselo a Electrolytic
1.Kuyeza ma elekitirodi 2. Kuyeza ma elekitirodi 3. Electrolytic electrode 4. Electrolytic electrode lead 5. Ion filter membrane 6. Kuyanika pulagi yamagalasi a chubu 7. Drying chubu 8. Allochroic silicagel (drying agent) 9. Chitsanzo cholowera 10. Stirrer 10. Stirre . Anode chipinda 12. Cathode chipinda 13. Electrolytic cell galasi kugaya pulagi
●Ikani tinthu tating'ono ta buluu ta silikoni (chowumitsa) mu chubu chowumitsa (7 mumkuyu 4-2).
Zindikirani: Chitoliro cha chubu chowumitsa chiyenera kukhala ndi mpweya wokwanira ndipo sichikhoza kusindikizidwa kwathunthu, mwinamwake ndizosavuta kuyambitsa zoopsa!
●Lowetsani tani ya silikoni yoyera yamkaka mu tambala ndikumapukuta mofanana ndi zomangira (onani mkuyu 4-4).
Chithunzi 4-4 jakisoni pulagi kujambula msonkhano
●Mosamala ikani chosonkhezera mubotolo la electrolytic kudzera polowera pachitsanzo.
●Ganizirani mogawana mafuta osanjikiza pamagetsi oyezera, ma elekitirodi a electrolytic, chubu chowumitsa chachipinda cha cathode, ndi doko lopukutira tambala.Mukayika zinthu zomwe zili pamwambazi mu botolo la electrolytic, tembenuzani pang'onopang'ono kuti likhale losindikizidwa bwino.
●Pafupifupi 120-150 ml ya electrolyte imalowetsedwa mu chipinda cha anode cha cell electrolytic kuchokera ku doko lotsekera ma cell a electrolytic ndi funnel yoyera ndi youma (kapena kugwiritsa ntchito chosinthira madzi), komanso kubayidwa muchipinda cha anode cha cell electrolytic electrolytic elekitirodi kusindikiza doko ndi funnel (kapena ntchito madzi osintha), kuti electrolyte mlingo mkati cathode chipinda ndi anode chipinda ali chimodzimodzi.Mukamaliza, pulagi yogaya magalasi ya cell electrolytic imakutidwa mofanana ndi wosanjikiza wa vacuum grease ndikuyika pamalo ofananirako, imazunguliridwa mofatsa kuti ikhale yosindikizidwa bwino.
Chidziwitso: Ntchito yonyamula ma electrolyte pamwambapa iyenera kuchitidwa pamalo olowera mpweya wabwino.Osapumira kapena kukhudza ma reagents ndi dzanja.Ngati yakhudza khungu, muzimutsuka ndi madzi.
Mukamaliza masitepe omwe ali pamwambawa, ikani selo la electrolytic mu chothandizira cha electrolytic cell (9 mu mkuyu 4-1), ikani waya wolumikizira ma elekitirodi a electrolytic ndi pulagi ya lotus ndi waya woyezera ma elekitirodi olumikizira mu mawonekedwe a electrolytic electrode ( 7 mkuyu. 4-1).) ndi mawonekedwe a electrode yoyezera (8 mu Fig.4-1).
The reagent yankho ndi chisakanizo cha ayodini, pyridine wodzazidwa ndi sulfure dioxide ndi methanol.Mfundo ya Karl-Fischer reagent ndi madzi ndi: kutengera kukhalapo kwa madzi, ayodini imachepetsedwa ndi sulfure dioxide, ndipo pamaso pa pyridine ndi methanol, pyridine hydroiodide ndi methyl hydrogen hydrogen pyridine amapangidwa.Reaction formula ndi:
H20+I2+SO2+3C5H5N → 2C5H5N·HI+C5H5N·SO3 …………(1)
C5H5N·SO3+CH3OH → C5H5N·HSO4CH3 ……………………(2)
Pa electrolysis ndondomeko, electrode anachita motere:
Anode: 2I- - 2e → I2 ..........................................(3)
Cathode: 2H+ + 2e → H2↑..........................................(4)
Iodine yopangidwa ndi anode imachita ndi madzi kupanga hydroiodic acid mpaka kutha kwamadzi onse, ndipo kutha kwake kumawonetsedwa ndi gawo lozindikira lomwe limapangidwa ndi ma electrode a platinamu.Malinga ndi lamulo la Faraday la electrolysis, chiwerengero cha mamolekyu a ayodini omwe akugwira nawo ntchito ndi ofanana ndi chiwerengero cha mamolekyu a madzi, omwe amafanana ndi kuchuluka kwa magetsi.Kuchuluka kwa madzi ndi mtengo wake uli ndi equation iyi:
W=Q/10.722 ……………………………………………… (5)
W--chinyezi zachitsanzo Unit: ug
Q-electrolysis kuchuluka kwa charger yamagetsi Unit: mC
Chidacho chimatenga LCD yazithunzi zazikulu, ndipo kuchuluka kwa chidziwitso chomwe chikhoza kuwonetsedwa pazithunzi zonse ndizolemera, zomwe zimachepetsa chiwerengero cha zowonetsera.Ndi mabatani okhudza, ntchito za mabatani zimafotokozedwa momveka bwino, zosavuta kugwiritsa ntchito.
Chidachi chimagawidwa m'mawonekedwe 5:
●Boot yolandirira skrini;
●Screen yokhazikitsa nthawi;
●Mbiri yakale yotchinga;
●Tsamba loyesera lachitsanzo;
●Chophimba chotsatira choyezera;
1. Boot Welcome Screen
Lumikizani chingwe chamagetsi ndi kuyatsa chosinthira magetsi.Chithunzi cha LCD 6-1 chikuwonetsedwa:
2.Time Kukhazikitsa Screen
Dinani batani la "Nthawi" mu mawonekedwe a Chithunzi 6-1, ndipo chophimba cha LCD chidzawonetsedwa monga momwe chithunzi 6-2:
Mu mawonekedwe awa, kanikizani gawo lanthawi kapena tsiku kwa masekondi atatu kuti muyike kapena kuwerengera nthawi ndi tsiku.
Presskiyi kuti mubwerere ku mawonekedwe a boot.
3. Historical Data Screen
Dinani batani la "Data" pachithunzi cha Chithunzi 6-1, ndipo chophimba cha LCD chidzawonetsedwa monga momwe chithunzi 6-3:
Press kiyi yosintha masamba.
PressKiyi yochotsa zomwe zilipo.
Presskiyi yosindikiza zomwe zilipo.
Presskiyi kuti mubwerere ku mawonekedwe a boot.
4. Chitsanzo Mayeso Screen
Dinani batani la "Mayeso" pazenera la Chithunzi 6-1, chophimba cha LCD chidzawonetsedwa monga momwe zilili pansipa:
Ngati electrolyte mu cell electrolytic yasinthidwa kumene, zomwe zilipo zikuwonetsa "Reagent pa ayodini, chonde lembani madzi".Pambuyo pobaya madzi pang'onopang'ono m'chipinda cha anode ndi 50ul sampler mpaka electrolyte itasanduka chikasu, zomwe zilipo zikuwonetsa "Chonde dikirani", ndipo chidacho chidzakhazikika.
Ngati electrolyte mu cell electrolytic yakhala ikugwiritsidwa ntchito, zomwe zikuchitika zikuwonetsa "Chonde dikirani", ndipo chidacho chidzakhazikika.
Pre-conditioning akuyamba, mwachitsanzo titration chotengera si zouma."Chonde dikirani" iwonetsa, chida chosinthira madzi owonjezera.
PressKiyi yosankha zinthu.
PressKiyi yoyambira mayeso.
Presskiyi kuti mubwerere ku mawonekedwe a boot
4.1 Mu mawonekedwe awa, akanikizire "Ikani" kiyi, ikani yogwira mtima liwiro ndi Ext.nthawi.
Chithunzi 6-5
Dinani liwiro loyambitsa (gawo lachiwerengero) kuti muyike kuthamanga kwa chipangizocho.Dinani Ext.nthawi (gawo lachiwerengero) kuti muyike nthawi yochedwa ya mapeto a mayeso.
Kuthamangitsa liwiro: Pamene mamasukidwe akayendedwe a chitsanzo choyesedwa ndi chachikulu, liwiro oyambitsa akhoza bwino kuchuluka.Kupanda thovu mu oyambitsa electrolyte.
Zowonjezera.Nthawi: Pakafunika kukulitsa nthawi yoyesera yachitsanzo, monga kusasungunuka kwachitsanzo ndi electrolyte kapena madzi oyesera a gasi, nthawi yoyesera imatha kukulitsidwa moyenera.(Zindikirani: Pamene Ext. nthawi yakhazikitsidwa kwa mphindi 0, kuyesedwa kumatsirizidwa pambuyo poti liwiro la electrolysis la chipangizocho likhale lokhazikika. Pamene Ext. nthawi yakhazikitsidwa kwa mphindi 5, kuyesa kumapitirira kwa mphindi 5 pambuyo pa kuthamanga kwa electrolysis. chida chili chokhazikika)
4.2 Chiwongola dzanja chikamalizidwa, zomwe zilipo zikuwonetsa "Presskiyi yoyezera". Panthawiyi, chidacho chikhoza kuyesedwa kapena chitsanzo chikhoza kuyesedwa mwachindunji.
Kuti muyese chidacho, gwiritsani ntchito 0.5ul sampler kuti mutenge 0.1ul ya madzi, dinani batani la "Yambani", ndikuyibaya mu electrolyte kudzera muzolowera.Ngati zotsatira zomaliza za mayeso zili pakati pa 97-103ug (sampler yotumizidwa kunja), zimatsimikizira kuti chidacho chili bwino ndipo chitsanzocho chikhoza kuyesedwa.(Zotsatira zoyesa za sampler zapakhomo zili pakati pa 90-110ug, zomwe zimatsimikizira kuti chidacho chili bwino).
4.3 Chitsanzo cha Titration
Chidacho chikakhala moyenera (kapena kuwerengeredwa), zomwe zikuchitika pano ndi "Titrating", ndiye kuti chitsanzocho chikhoza kusinthidwa.
Tengani zitsanzo zoyenerera, dinani batani la "Start", lowetsani chitsanzocho mu electrolyte kudzera muzolowera zachitsanzo, ndipo chidacho chidzayesa mpaka kumapeto.
Zindikirani: Voliyumu yoyeserera imachepetsedwa moyenerera kapena kuonjezeredwa molingana ndi kuchuluka kwa madzi achitsanzocho.Zitsanzo zochepa zitha kutengedwa ndi 50ul sampler kuti mukayesedwe.Ngati mtengo wamadzi woyezedwa ndi wocheperako, kuchuluka kwa jekeseni kumatha kuonjezedwa moyenera;Ngati mtengo wamadzi oyezedwa ndi waukulu, kuchuluka kwa jekeseni kumatha kuchepetsedwa moyenera.Ndikoyenera kusunga zotsatira zomaliza zoyesa madzi pakati pa makumi a micrograms ndi mazana a micrograms.Mafuta a Transformer ndi mafuta a turbine amatha kubayidwa mwachindunji 1000ul.
5. Zotsatira Zoyezera
Kuyesa kwachitsanzo kukamalizidwa, chilinganizo chowerengera chikhoza kusinthidwa momwe chimafunikira, ndipo nambala yomwe ili kumanja kwa chilinganizocho imatha kusintha pakati pa 1-5.(zogwirizana ndi ppm, mg/L ndi% motsatana)
Muyezo wodziwika bwino wa chida ichi ndi 0μg-100mg.Kuti mupeze zotsatira zolondola zoyezetsa, kuchuluka kwa jekeseni kuyenera kuwongoleredwa moyenera malinga ndi kuchuluka kwa chinyezi cha zitsanzo zoyeserera.
1. Chitsanzo chamadzimadzi
Kuyeza kwachitsanzo chamadzimadzi: zitsanzo zoyesedwa ziyenera kuchotsedwa ndi jekeseni wachitsanzo, kenako ndikubayidwa muchipinda cha anode cha cell electrolytic kudzera pa doko la jakisoni.Musanayike chitsanzo, singano iyenera kutsukidwa ndi pepala losefera.Ndipo nsonga ya singano iyenera kuyikidwa mu electrolyte popanda kukhudzana ndi inwall ya electrolytic cell ndi electrode pamene mayeso ayesedwa.
2. Chitsanzo cholimba
Zitsanzo zolimba zitha kukhala za ufa, tinthu tating'onoting'ono kapena totchinga chisokonezo (chidutswa chachikulu chiyenera kuswa).Evaporator yamadzi yoyenera idzasankhidwa ndikulumikizidwa ndi chida pamene kuyesa kuyesa kumakhala kovuta kusungunuka mu reagent.
Kutenga chitsanzo cholimba chomwe chingathe kusungunuka mu reagent monga chitsanzo kufotokozera jekeseni yolimba, motere:
Chithunzi 7-1
1) Injector yolimba yachitsanzo ikuwonetsedwa ngati chithunzi 7-1, iyeretseni ndi madzi ndikuumitsa bwino.
2) Tsitsani chivindikiro cha jekeseni wolimba, jekeseni chitsanzo choyesera, phimba chivindikiro ndikuyeza molondola.
3) Tsitsani tambala wa pulagi wa doko la jakisoni wa electrolytic cell, ikani jekeseni wachitsanzo mu doko la jakisoni molingana ndi mzere wonse womwe wawonetsedwa ngati chithunzi 7-2.Tembenuzani jekeseni yolimba yachitsanzo cha madigiri 180 chosonyezedwa ngati mzere wa madontho mu chithunzi 7-2, kupangitsa kuyesa kuyesa kutsika mpaka muyeso utatha.M'kati mwake, zitsanzo zolimba zoyesa sizingagwirizane ndi ma elekitirodi a electrolytic ndi electrode yoyezera.
Chithunzi 7-2
Yesani jekeseni wa chitsanzo ndi chivindikiro molondola kachiwiri pambuyo jekeseni.Ubwino wa zitsanzo ukhoza kuwerengedwa molingana ndi kusiyana pakati pa miyeso iwiri, yomwe ingagwiritsidwe ntchito kuwerengera chiŵerengero cha madzi.
3. Chitsanzo cha gasi
Kuti chinyontho cha gasi chitengedwe ndi reagent, cholumikizira chidzagwiritsidwa ntchito kuwongolera zitsanzo zoyeserera kuti zilowe mu cell ya electrolytic nthawi iliyonse. (onani chithunzi 7-3).Chinyezi mu zitsanzo zoyezera gasi chikayezedwa, reagent pafupifupi 150ml iyenera kubayidwa mu cell electrolytic kutsimikizira kuti chinyezi chitha kuyamwa mokwanira.Panthawi imodzimodziyo, kuthamanga kwa gasi kumayendetsedwa pa 500ml pa mphindi.pafupifupi.Ngati reagent imachepetsa mwachiwonekere pakuyeza, pafupifupi 20ml glycol iyenera kubayidwa ngati chowonjezera.(mankhwala ena reagent akhoza kuwonjezeredwa monga pa chitsanzo chenicheni choyezera.)
Chithunzi 7-3
A. Kusungirako
1. Khalani kutali ndi kuwala kwa dzuwa, ndipo kutentha kwa chipinda kuyenera kukhala mkati mwa 5℃~35℃.
2. Musayike ndikuyigwiritsa ntchito pansi pa chilengedwe chokhala ndi chinyezi chachikulu komanso kusinthasintha kwakukulu kwa magetsi.
3. Osayiyika ndikuyigwiritsa ntchito pansi pa chilengedwe ndi mpweya wowononga.
B. Kusintha kwa silicone pad
Pad ya silicone mu doko la jekeseni wa zitsanzo iyenera kusinthidwa panthawi yake chifukwa chakuti kugwiritsidwa ntchito kwa nthawi yaitali kumapangitsa kuti pinhole ikhale yosasunthika ndikulowetsa chinyezi, zomwe zidzakhudza kuyeza. (onani chithunzi 4-4)
1. Kusintha kwa allochroic silicagel
Allochroic silicagel mu chitoliro chowumitsa chiyenera kusinthidwa mtundu wake ukasanduka buluu wopepuka kuchokera ku buluu.Osayika ufa wa silicagel mu chitoliro chowumitsa mukasintha, apo ayi utsi wa cell electrolytic udzatsekedwa zomwe zimapangitsa kuti electrolysis ithe.
2. Kusamalira doko la electrolytic cell polishing
Sinthani doko lopukutira la cell electrolytic masiku 7-8 aliwonse.Ikakanika kusinthasintha mosavuta, ivaleni ndi vacuum grisi pang'ono ndikuyiyikanso, apo ayi ndizovuta kuyichotsa ngati nthawi yantchito ndi yayitali kwambiri.
Ngati ma elekitirodi sangatsitsidwe, chonde musatulutse mokakamiza.Panthawi imeneyi, kuviika lonse electrolytic selo mu madzi ofunda kwa maola 24-48 mosalekeza, ndiye ntchito.
3. Kuyeretsa khungu la electrolytic
Tsegulani mphete yonse ya botolo lagalasi la electrolytic cell.Chotsani botolo la cell electrolytic, chitoliro chowumitsa, pulagi yosindikiza ndi madzi.Yanikani mu uvuni (kutentha kwa uvuni ndi pafupifupi 80 ℃) mukatsuka, kenako muziziziritsa mwachilengedwe.Mowa wa ethyl ukhoza kugwiritsidwa ntchito kuyeretsa electrolusis electrode, pamene madzi ndi oletsedwa.Mukamaliza kuyeretsa, ziumeni ndi chowumitsira.
Zindikirani: Osayeretsa ma elekitirodi, monga momwe chithunzi 8-1 chikusonyezera
Chithunzi 8-1
C. Bweretsani Electrolyte
1. Chotsani ma elekitirodi a electrolytic, electrode yoyezera, chubu choyanika, pulagi ya jakisoni ndi zinthu zina pabotolo la selo la electrolytic.
2. Chotsani electrolyte kuti ilowe m'malo mwa botolo la electrolytic cell.
3. Yeretsani botolo la selo la electrolytic, electrolytic electrode ndi electrode yoyezera ndi ethanol mtheradi.
4. Yanikani botolo la selo loyeretsedwa la electrolytic, electrolytic electrode, etc. mu uvuni wosapitirira 50 ℃.
5. Thirani electrolyte yatsopano mu botolo la electrolytic cell, ndikutsanulira kuchuluka kwa pafupifupi 150ml (pakati pa mizere iwiri yopingasa yoyera ya botolo la electrolytic cell).
6. Ikani ma elekitirodi a electrolytic, electrode yoyezera, ndi pulagi yowuma ya chubu, ndi zina zotero, ndikutsanulira electrolyte yatsopano mu electrode electrode, kuchuluka kwake komwe kumatsanulidwa kumakhala kofanana ndi mlingo wamadzimadzi a electrolyte mu botolo la electrolytic cell.
7. Ikani mafuta osanjikiza a vacuum pamadoko onse opera a selo ya electrolytic (electrolytic electrode, electrode yoyezera, pulagi ya jekeseni, pulagi yopera magalasi).
8. Ikani botolo la selo losinthidwa la electrolytic mu botolo la electrolytic cell clamp ya chidacho, ndikusintha chidacho kuti chikhale chokhazikika.
9. The reagent yatsopano iyenera kukhala yofiira-bulauni komanso mu chikhalidwe cha ayodini.Gwiritsani ntchito jekeseni wa 50uL kuti mubaye madzi pafupifupi 50-100uL mpaka jekeseniyo itasanduka chikasu.
1. Palibe chiwonetsero
Chifukwa: Chingwe chamagetsi sichimalumikizidwa;chosinthira magetsi sichimalumikizana bwino.
Chithandizo: Lumikizani chingwe chamagetsi;sinthani chosinthira mphamvu.
2. Tsegulani dera la electrode yoyezera
Chifukwa: Elekitirodi yoyezera ndi pulagi ya chida sichilumikizidwa bwino;waya wolumikizira wathyoka.
Chithandizo: Lumikizani pulagi;sinthani chingwe.
3. Kuthamanga kwa electrolysis nthawi zonse kumakhala ziro panthawi ya electrolysis.
Chifukwa: Elekitirodi electrolytic ndi pulagi chida si kugwirizana bwino;waya wolumikizira wathyoka.
Chithandizo: Lumikizani pulagi;sinthani chingwe.
4. Calibration zotsatira za madzi oyera ndi ang'onoang'ono, pamene mayeso chitsanzo jekeseni, sangathe wapezeka ndi chida.
Chifukwa: Electrolyte imataya mphamvu.
Chithandizo: M'malo mwa electrolyte yatsopano.
5. Electrolytic ndondomeko sangathe kutha.
Chifukwa: Electrolyte imataya mphamvu.
Chithandizo: M'malo mwa electrolyte yatsopano.