GD6900 Capacitance ndi Dissipation Factor Tester
GD6900 imayesa mphamvu ndi dielectric loss factor (tgδ) ya zida zamagetsi zapamwamba kwambiri.Ndi mawonekedwe ophatikizika, mlatho woyeserera wotayika wa dielectric, magetsi osinthika osinthika, owonjezera thiransifoma ndi SF6 capacitor wamba.
Gwero lamphamvu lamagetsi limapangidwa ndi inverter yamkati ya chidacho, chomwe chimagwiritsidwa ntchito poyesa chinthu choyezera pambuyo pakukweza kwa thiransifoma.Ngati muwonjezera kapu yotchinjiriza ndi chipangizo chowongolera kutentha, kutayika kwa dielectric kwa mafuta oteteza kungathe kuyesedwa.
●Kukhudza LCD skrini chiwonetsero.
●Muyeso wa kukana kwa insulation: Gawo loyeserera lokhazikika la insulation resistance (IR), mayamwidwe ratio (DAR), polarization index (PI) test.
●Kalendala chip ndi chosungira chachikulu mkati.Sungani zotsatira zoyesa malinga ndi dongosolo la nthawi, onani mbiri yakale ndikusindikiza zotsatira.
●Deta ya chida ikhoza kutumizidwa kunja kudzera pa U disk.
●Njira zingapo zoyesera, zokhala ndi ma voliyumu apamwamba, kunja kwamagetsi apamwamba, muyezo wamkati, muyezo wakunja, GST/UST, kudzisangalatsa.Kuyesa kwamphamvu kwamagetsi (kuposa 10kV) kutayika kwa dielectric kumatha kupangidwa ngati mulingo wakunja kunja kwa voteji yayikulu.
●Yesani CVT yosindikizidwa kwathunthu (Capacitive Voltage Transformer) C1 ndi C2 dielectric kutaya ndi mphamvu nthawi imodzi.Yesaninso chiŵerengero cha kusintha kwa CVT ndi kusiyana kwa ngodya yamagetsi.
●Kutayika kwa dielectric ndi capacitance value ya C0 kumapeto kwa CVT kungayesedwe pogwiritsa ntchito njira yotchinga kumbuyo.
●Liwiro lachitsanzo chizindikiro.Inverter ndi sampuli zozungulira mkati zimayendetsedwa ndi digito.Kutulutsa mphamvu kumasinthidwa mosalekeza.
●Kuyeza kwa LCR automatic.Inductance, capacitance, kukana, ngodya imatha kuyesedwa ndikuwonetsedwa.
●Kutetezedwa kangapo kwa kusinthasintha kwamagetsi, kutulutsa kwafupipafupi, kupitilira-voltage, kupitilira apo, kutentha, kutetezedwa komanso kudalirika.Pakadali pano, ili ndi ntchito yoyesa pansi, kuti kukwera kwamagetsi sikuloledwa pazida zopanda pansi.
●Palibe chifukwa chochotsa chiwongolero cha HV kuyeza kutayika kwa dielectric ndi capacitance ya CVT.
●Mafupipafupi amatha kusinthidwa kukhala 50Hz, 47.5Hz/52.5Hz, 45Hz/55Hz, 60Hz, 57.5Hz/62.5Hz, 55Hz/65Hz.
Mkhalidwe wogwirira ntchito | -15 ℃ ~ 40 ℃ | RH<80% | ||||||
Mfundo yotsutsana ndi kusokoneza | Kutembenuka pafupipafupi | |||||||
Magetsi | AC 220V±10%,50/60Hz | Jenereta angagwiritsidwe ntchito. | ||||||
Kutulutsa kwakukulu kwamagetsi | AC 0.5KV~10KV (12kV ngati mukufuna) | Pafupifupi 0.1 kV | ||||||
Kulondola | 2% | |||||||
Max.panopa | 200mA | |||||||
Mphamvu | 2000 VA | |||||||
Kudzisangalatsa mphamvu | AC 0V~50V/15A | 50Hz, 60Hz pafupipafupi 45HZ/55HZ 47.5HZ/52.5HZ 55HZ/65HZ 57.5HZ/62.5HZ Makina apawiri pafupipafupi | ||||||
Kusamvana | tgδ: 0.001% | Cx: 0.001pF | ||||||
Kulondola | ∆tgδ: ±(kuwerenga*1.0%+0.040%) | |||||||
∆C x: ±(kuwerenga*1.0%+1.00PF) | ||||||||
Muyezo osiyanasiyana | tg ndi | Popanda malire | ||||||
C x | 15pF <Cx <300nF | |||||||
10 kV | Cx <60 nF | |||||||
5 kV pa | Cx <150 nF | |||||||
1 kV pa | Cx <300 nF | |||||||
CVT mayeso | Cx <300 nF | |||||||
Muyezo wa LCR | L>20H(2kV) | R>10KΩ(2kV) | ||||||
Kulondola kwa kuyeza kwa LCR | 0.1% | Kusamvana | 0.01 | |||||
Chiwerengero cha CVT | 10-10000 | |||||||
Kulondola kwa chiŵerengero cha CVT | 0.1% | |||||||
Kusintha kwa chiwerengero cha CVT | 0.01 | |||||||
Insulation resistance | DC high voltage | 5-10 kV | Kulondola | ±(2%*Kuwerenga +10V) | ||||
Kulondola kwa IR | 100kΩ-1000GΩ: <5% (voteji yoyesa osachepera 250V) 100GΩ-1000GΩ: 10% (voteji yoyesera osachepera 10000V) | |||||||
Dimension | Chigawo chachikulu: 350(L)×270(W)×270(H) | |||||||
Bokosi lowonjezera: 350(L)×270(W)×160(H) | ||||||||
Kukhoza kukumbukira | Magulu 200, USB flash disk yosungirako imathandizidwa. | |||||||
Kulemera | Gawo lalikulu: 22.75Kg | |||||||
Bokosi lowonjezera: 5.25Kg |
Woyesa | 1 seti |
Chalk kesi | 1 chidutswa |
Mayeso ofiira a HV amatsogolera 8m | 1 chidutswa |
Mayeso akuda a LV amatsogolera 8m | 1 chidutswa |
Chingwe champhamvu 1.8m | 1 chidutswa |
CVT Kulumikiza kutsogolo 5m | 1 chidutswa |
Kutalika kwa 2m | 1 chidutswa |
Sindikizani pepala | 2 mitu |
10A fuse | 3 zidutswa |
Buku Logwiritsa Ntchito | 1 kopi |
Lipoti la Mayeso a Factory | 1 kopi |
Zosankha zowonjezera | |
Mafuta Oyesera cell | 1 pc |
Wowongolera kutentha | 1 pc |
Ndi cell test cell ndi chowongolera chotenthetsera, wogwiritsa ntchito amatha kuyesa tan delta yamadzi a dielectric monga mafuta osinthira.
Wowongolera kutentha
Chombo cha mafuta